Tuff TT
Tuff TT Implants share the same three-thread zone concept as Tuff implants. The Tuff TT’s converging coronal shape allows platform switch technology to reduce crestal bone loss.
Material: Titanium (Ti6Al4V ELI)
Treatment: RBM
10

 

13

12(1)

* The proposed procedure is only a recommendation and should not replace the doctor’s judgment.
The implants may be placed in immediate function when good primary stability (above 35 Ncm) has been achieved and with appropriate occlusal loading.
* Implant Carrier removal After the osteotomy preparation, the implant should be inserted with the aid of the implant carrier.
The implant should be initially stabilized by a few threads and then the carrier should be removed. Farther insertion of the implant will be done with appropriate tool.
* Implant hexagon During implant insertion, the hexagon of the implant, should be located with a straight part of the hexagon toward the angulation needed, in order
to provide adequate rehabilitation.